mT-Jump - Rapid Mixing Temperature Jump System

· Hot and cold jumps
· T-changes as high as several ten of degrees
· Low volume consumption
· Millisecond resolution

Principle
This new T-jump instrument achieves temperature changes by mixing two solutions of different initial temperatures T1 and T2. The final temperature of the mixture (T3) is calculated from the initial temperatures T1 and T2 and the mixing ratio of the two solutions.

Three thermoelectric Peltier elements are used to control the initial temperatures of the two solutions and that of the observation cell after mixing. The mT-jump accessory is installed on a Bio-Logic stopped-flow basis.

The system is fully controlled from the Bio-Kine32 software: the user defines the mixing ratio for the reaction and the amplitude of the temperature jump.
The temperature of each Peltier element is then automatically adjusted. Automatic accumulation of shots is possible to optimize the signal to noise ratio.

The mT-jump system is equipped with a standard Bio-Logic stopped-flow observation cell. It is compatible with all modes of observations such as circular dichroism, fluorescence, absorbance, and fluorescence anisotropy.

A complementary instrument to classical T-jump

The aim of classical T-jump instruments is to record fast heat induced relaxations of reactions. This is normally achieved by applying fast electric current pulses or more recently infrared laser pulses to the sample. Time resolution of classical T-jump can reach a few μs but the kinetics cannot be recorded for more than a few ms due to temperature dissipation, which could be a limitation in some applications. Most classical temperature jump devices are limited to 5°C and are unable to perform cooling jumps.

Bio-Logic’s mT-jump is based on Stopped-Flow technology and thus offers millisecond resolution. The amplitude of the jump is entirely controlled by the user (up to 40°C jump) and the jump can be in both directions (cooling or heating jump) which offers new investigation domains, for example: refolding studies. Protein folding and refolding can easily be studied without adding denaturant to the protein! Detection with classical T-jump instrumentation is also limited to techniques such as absorbance and fluorescence whereas mT-jump is compatible with all optical methods (including circular dichroism, FT-IR, fluorescence anisotropy…)

Specifications

Temperature control specifications
? Full software control through USB connection.
? Temperature range : +5°C / +90°C
? Precision of temperature reading: 0.01°C
? Maximum temperature jump : +/- 40°C
? Temperature stability for S1 and S2 in stationary state: < +/- 0.2°C at 80°C after 15 minutes of equilibrium.
? Temperature stability in the cuvette after the T-jump: variation < 1% per 30s
? Precision of the T-jump: defined by user calibration (down to +/- 0.1°C)
? Temperature probe is included.

Mixing specifications
? Compatible with all SFM.
? Mixing Ratio from 1/1 to 1/4.
? Sample consumption per shot : 45 μl to 220 μl.
? Dead time: depends on cuvette, stopped-flow model and mixing ratio (0.7ms using SFM-2000 series and FC-08 cuvette)
? compatible with MOS-200, MOS-450/AF-CD, Diode Array detector and external devices.
? optional FT-IR cell.

Copyright Dalian Huayang analytical instruments Co., Ltd.All right reserved. Preparation of Liao ICP
友情鏈接: link1 link2 link3 link4 link5 link8 link9
主站蜘蛛池模板: 亚洲国产日韩女人aaaaaa毛片在线 | 中文字幕乱码在线 | 国产专区一区 | 中文字幕91在线 | 久久99精品久久久久久首页 | 久草热草 | 亚洲国产精品一区二区三区久久 | 成人爱爱免费视频 | 91精品专区 | 久久精品香蕉 | 精品福利三区3d卡通动漫 | 香蕉尹人综合精品 | 日本不卡视频在线播放 | 国产成人艳妇在线观看 | 久久欧美精品欧美久久欧美 | 国产在线不卡午夜精品2021 | 成人午夜精品久久不卡 | 四虎精品免费久久 | 免费国产成高清人在线视频 | 欧美韩国日本在线 | 精品国产免费一区二区三区 | 天天操天天操天天操香蕉 | 日日操夜夜添 | 国产精品久久久久毛片真精品 | 久久综合久久自在自线精品自 | 欧美日韩大尺码免费专区 | 暖暖激情视频 | 国产青青在线视频 | 日韩午夜大片 | 最新九九精品 | 国产成人精品亚洲 | 久久久久色 | 校花超短裙暴露调教在线观看 | 亚洲跨种族黑人xxx 亚洲另类 专区 欧美 制服 | 天天操天天摸天天干 | 欧美男同gay | 中文字幕一区精品欧美 | 亚洲成人精品久久 | 午夜免费69性视频爽爽爽 | 日本视频www | 欧美成人看片黄a免费 |